Brand Name: | XL |
Model Number: | XL-803J |
Certification: | CE |
Place of Origin: | China |
MOQ: | negotiate |
Price: | negotiate |
Payment Terms: | L/C, D/A, D/P, T/T, Western Union, MoneyGram |
Supply Ability: | 800 pcs per month |
Delivery Time: | 3~4 weeks |
Packaging Details: | external packing:carton or non wooden box;cushion packing:EPE foam |
Name: | Multifunction Electrical Calibrator |
AC voltage Adjustment fineness: | 0.02% RG |
AC voltage Accuracy: | better than ±0.1% RG |
AC voltage Stability: | ±0.02% RG/1min |
AC voltage Output power rate: | 15VA per phase |
AC voltage Output range: | 0V~264V |
AC voltageGear setting: | 0V~130V, 130V~264V Automatic shifting |
AC voltage Temperature drift: | ±16PPM/°C |
High Stability AC Voltage Calibrator , Multifunction Tester Calibration Energy Meter
1 . Overview
The XL-803J three-phase AC Voltage Calibrator programmable standard
power source is a new generation of high-precision standard power
source based on 1.2 G MAC DSP, large-scale FPGA, high-speed and
high-precision DA and high-fidelity power amplifier. Applicable to
electric power system, electric power, remote control, dispatching
and other power departments and enterprises that need measurement,
inspection and high-precision standard signal sources. It is also
suitable for other occasions where high-precision standard signal
sources are required for measurement and inspection.
XL-803J can output high-precision voltage and current with
adjustable frequency, phase and amplitude at power frequency
(40Hz~65Hz). It is a highly accurate adjustable voltage and current
standard source. The XL-803J can output very pure sinusoidal
voltage current with a distortion of no more than 0.2%, typically
0.05%. The XL-803J's voltage and current output has very high
output stability, typically 0.02% RD. Therefore, it is very
suitable for work situations requiring high-precision inspection
and calibration, such as the measurement department for the
detection of various voltage, current, power and other electrical
parameter meters.
2. Main features
1) It is possible to output a pure sinusoidal power signal with a
distortion of 0.05% (typical).
2) Each harmonic output can be superimposed on the fundamental
wave.
3) The frequency output can be adjusted from 40Hz to 65Hz, the
resolution is 0.001Hz, and the accuracy is 0.004Hz.
4) The AB phase is a frequency reference, and the C phase is a
separate frequency reference, so it can be phase-separated.
5) The phase 0~360 degrees can be adjusted arbitrarily, which can
easily simulate various power supply situations and even reverse
power transmission.
6) Strong load capacity, can carry capacitive, inductive, resistive
load or composite type load, and the load regulation rate is better
than 0.02% RG.
7) Excellent temperature stability, the core device is a military
grade product with a temperature coefficient as small as 1PPM,
which can guarantee the output precision in the outdoor temperature
environment.
8) It is powerful and flexible with 32-bit MCU+DSP processor.
9) The waveform of the power frequency is up to 50,000 points per
week, and the internal signal output does not need a filter for
smooth filtering, which ensures the accurate output of the
waveform, so that the system can output accurate harmonics, and the
system has excellent harmonic distortion. index.
10) It can be easily connected to the PC through an RS232 to expand
other functions.
11) Perfect overcurrent, overvoltage, overheat, short circuit, open
circuit, overload protection.
12) Hardware PID, the response is extremely fast, and the change of
load will not cause the fluctuation of the output.
13) 320*240 liquid crystal display, Chinese interface, easy to
operate.
14) Open communication protocol to facilitate secondary development
(factory automatic verification of RTU/FTU/power management
terminal/common measurement terminal).
15) Can carry a pure capacitive load. The energy meter can be
checked in conjunction with the PC software.
3. Main functions
Main function Description
AC standard source output can output power frequency (40Hz ~ 65Hz)
frequency, phase and amplitude adjustable high-precision voltage
and current, convenient for power workers to develop, verify
Transmitter verification for voltage, current, power verification
Communication function Used to communicate with PC and other main
control modules, only supports "Programmable Power Interface
Protocol"
Power check function for verifying energy pulse input
Split-phase frequency output The A and B phase frequencies of the
device are uniformly adjustable, and the C-phase frequency can be
independently adjusted, which is convenient for users who need two
frequencies, such as the synchronous device in power protection.
User-defined functions User-definable function output, but requires
functions that do not contain DC components
Secondary development Users can refer to the "Programming Power
Interface Protocol" to develop various functions required through
the communication port.
Software control function Equipped with XL-1000 series operating
software, as long as the device power is turned on and the
communication line is connected, it can be controlled freely on the
PC, which is convenient and quick.
Local function with 320*240 LCD display and 40 buttons for local
operation
Alarm function When the device is abnormal, the system will
automatically start the protection program, accompanied by a sound
or interface pop-up prompt. At this time, the user should power off
and find the cause of the fault and eliminate it, such as voltage
to ground short circuit, current open circuit output.
4. Main Technical Indicators
4.1. AC voltage output
Name: | Multifunction Electrical Calibrator | AC Voltage Adjustment Fineness: | 0.02% RG |
---|---|---|---|
AC Voltage Accuracy: | Better Than ±0.1% RG | AC Voltage Stability: | ±0.02% RG/1min |
AC Voltage Output Power Rate: | 15VA Per Phase | AC Voltage Output Range: | 0V~264V |
AC VoltageGear Setting: | 0V~130V, 130V~264V Automatic Shifting | AC Voltage Temperature Drift: | ±16PPM/°C |
4.2. AC current output
Name: | Multifunction Electrical Calibrator | AC Current Adjustment Fineness: | 0.02% RG |
---|---|---|---|
AC Current Accuracy: | Better Than ±0.1% RG | AC Current Stability: | ±0.02% RG/1min |
AC Current Output Power Rate: | 15VA Per Phase | AC Current Output Range: | 0A ~ 6A |
AC Current Gear Setting: | 0A~1A, 1A~6A Automatic Shifting | AC Current Temperature Drift: | ±16PPM/°C |
![]() |